Salina Nova

Inhalation for professional use

Electrical data:

  • Rated voltage and frequency: 230 VAC / 50Hz
  • Power consumption: max. 250 VA
  • Fuse 230V (10): T0.63A H, 250V (T1.0A H, 250V with ADV option)
  • Classification: Protection class I, type B
  • Operating mode: Continuous operation
  • IP degree of protection: IP 30 (IP 20 when equipped with manual nebulizer (AHV option) or nasopharyngeal douche (NRD option))
  • Schalldruckpegel: < 58 dB(A)
  • Firmware version: see type plate

Performance data:

  • Nominal operating pressure: 200 kPa (2 bar)
  • min. Connection pressure: 220 kPa (2.2 bar)
  • max. connection pressure: 600 kPa (6 bar)
  • Air volume: 20 l/min. per nozzle
  • Aerosol delivery: 0.2 - 2.2 ml /min* (depending on the combination of activated nebulizer units)
  • Aerosol temperature normal operation: 35 °C
  • Aerosol temperature: max. 45 °C
  • Flow rate: 60 l/min max.
  • ADV flow rate switched on: 80 l/min max.
  • Outlet pressure (NRD): 200 kPa (2 bar) max.
  • Nasopharyngeal douche (NRD) flow rate: 10 l/min max.
  • Manual nebulizer outlet pressure (AHV): 200 kPa (2 bar)
  • Manual nebulizer flow rate (AHV): 10 l/min max.

*Performance data does not apply to medications that are in suspension or highly viscous.

Operating conditions:

  • Operating temperature: 15 to 30°C
  • Humidity: 0 to 95% RH, non-condensing
  • Air pressure: 800 to 1060 hPa
  • Working height: max. 2000 m above sea level

Transport and storage conditions:

  • Storage temperature: -20 to 70°C
  • Humidity: 0 to 95% RH
  • Air pressure: 800 to 1060 hPa

Dimensions and weights:

  • Dimensions: 450 x 470 x 420 mm (W x H x D)
  • Weight: 16.0 kg max.

Subject to technical changes!