Oxygen ionization

Oxygen activation at the highest level

Electrical data:

  • Rated voltage, frequency: 230 VAC / 50Hz
  • Power consumption: max. 350 VA
  • Fuse mains side: 2 x T3,15A H 250V
  • Fuse internal: T1,0A L 250V
  • Schalldruckpegel: < 48 dB(A)
  • Protection class: Protection class I
  • Application part (breathing mask): Type B
  • Operating mode: Continuous operation
  • IP classification: IP 21

Dimensions:

  • Dimensions: 535 x 203 x 520 mm (H x W x D) without rollers
  • Weight: 19.9 kg

Storage and transport conditions:

  • Storage temperature: -25 to +70°C
  • Humidity: 15 to 93% RH, non-condensing
  • Air pressure: 700 to 1060 hPa

Operating conditions:

  • Operating temperature: -5° to +40°C
  • Humidity: 15 to 93% RH, non-condensing
  • Air pressure: 700 to 1060 hPa
  • Working altitude: max. 3000 m above sea level

Performance data:

  • O2 concentration at 0.1 to 5 l/min: 93 % +/- 3 % 1)
  • O2 concentration at 5 to 6 l/min: 90 % +/- 3 % 1)
  • Volume flow: 0.1 - 6 l/min 1)
  • max. recommended flow rate: 6 l/min
  • max. outlet pressure: 55 kPa (normal condition), 275 kPa (initial fault)
  • Ion concentration: 5 million/cm3 (+/- 20%) at 4 l/min

1) O2 concentration and flow rate data apply over the full range of operating conditions specified above.

2) Flow rate data at: Room temperature and air pressure, dry air.

Software version:

  • Oxygen concentrator:
    Software version is briefly displayed on the LC display of the oxygen concentrator after it is switched on.
  • Ionizing unit:
    Software version is shown on the display of the ionizing unit for 10 seconds after switching on.

Subject to technical changes!